Ventilation and Control
· natural convection
· continuous adjustment of pre-heated fresh air admixture
· vent connection with restrictor flap
· fuzzy-supported PID microprocessor controller with integrated autodiagnostic system with fault indication
· solid state switching unit
· 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
· digital 7-day-programme-timer with real time clock, precise minute setting, for one set
value or ramp operation in combination with:
· integrated digital timer for tempering profile with max. 4 segments, each segment adjustable from 1 min to 999 hours:
DELAYED ON
HEATING UP
HOLD or HOLD set-temperature-dependent
with guaranteed holding time
COOLING DOWN
· "Loop" function to repeat profile 1-99 times or
endless
· digital display (LED) of all set parameters, such as temperature, weekdays, time, programme status and set-up values
· resolution of display for set value 0,5 °C; for actual value 0,1 °C below 99,9 °C, 0,5 °C
above 100 °C
· long-term logging (ring store) of all relevant
data, GLP-conforming as data logger - 1024
kB
· serial interface RS 232 including MEMMERT
Software "Celsius" for programming and
documentation
· incl. works calibration certificate for +160 °C
Telp. : 0878 8545 8787